![Dräger Atlan A300 Instructions For Use Manual Download Page 235](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728235.webp)
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
235
Reprocessing
On the product variant with integrated patient-gas measurement module, the
sealing cap must be plugged into the holder.
13.8.6
Removing the breathing system
To prevent accidental penetration of soda lime into the breathing system, make
sure that the reusable CO
2
absorber has been removed.
Prerequisite: The breathing system cover has been removed.
1. To unlock, turn the 3 levers (1) counterclockwise by approximately 120°.
2.
CAUTION
Risk of injury due to breathing system warmer
When the breathing system warmer is switched on, the bottom side of the
breathing system and the heating plate beneath it can become very hot.
►
Allow the breathing system to cool off before removing.
41
134
37063