Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
!
100
Central O
2
supply high
The central supply pressure
is high. The gas delivery
might fail.
Check central supply.
!!! 210
Central O
2
supply low
Central supply pressure is
low.
Check central O
2
supply or
use cylinder.
Use "ALARM RESET" to
downgrade alarm priority.
!
255
Central O
2
supply low
Central supply pressure is
low.
Check central supply.
!
100
CO
2
absorber expired
Accessory has been used
too long.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
acknowledge alarm.
!
255
CO
2
measurement failed
There is electromagnetic
interference.
Check for electromagnetic
radiation in the vicinity.
There is an internal failure.
Use an alternative gas mea-
surement system.
If the problem persists, call
Dräger.
!
255
CO
2
sensor accuracy low
The sensor has not yet
warmed up.
Wait for the automatic cali-
bration.
The ambient air used for cali-
brating the sensor was
impure.
Position the device in an
environment with clean ambi-
ent air.
!
0
Communication failure
The network connection
could not be established.
Re-establish the connection.
Use "ALARM RESET" to
acknowledge alarm.
Check the connections
from/to Connectivity Con-
verter CC300.
A network certificate has
expired or Connectivity Con-
verter CC300 is faulty.
Call Dräger.
In the network, there is
another anesthesia machine
with the same integrated sys-
tem ID.
Check the integrated system
ID.
!! 135
Cooling fan failure
An internal fan for evacuating
gases is faulty.
To prevent potential dam-
age, turn off the device at
your earliest convenience.
Increased risk of fire.
Call Dräger.
Priority
Alarm
Cause
Remedy