![Dräger Atlan A300 Instructions For Use Manual Download Page 276](http://html.mh-extra.com/html/drager/atlan-a300/atlan-a300_instructions-for-use-manual_2526728276.webp)
276
Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Resolution of displayed value
1 hPa (cmH
2
O)
Pressure gauge for indicating the pressure
in the internal breathing system
Range
–20 to +80 hPa (cmH
2
O)
Accuracy
±5 % of the measured value or ±2 hPa
(cmH
2
O) (the larger value applies)
Resolution of displayed value
5 hPa (cmH
2
O)
Volume
The measured volume values displayed
already take account of the compliance of the
breathing circuit. As soon as CO
2
respiratory
phases are detected, the sample gas flow for
the patient-gas measurement module is addi-
tionally taken into account.
Tidal volume
Expiratory
VT
Range
0 to 2500 mL
Accuracy for gas measurement with PGM
±10 % of the measured value or ±15 mL (the
larger value applies)
Accuracy without PGM
±20 % of the measured value or ±40 mL (the
larger value applies)
Resolution of displayed value
1 mL
Difference between inspiratory and expira-
tory tidal volumes
∆
VT
Range
0 to 2500 mL
Accuracy for gas measurement with PGM
±20 % of the measured value or ±30 mL (the
larger value applies)
Accuracy without PGM
±40 % of the measured value or ±80 mL (the
larger value applies)
Resolution of displayed value
1 mL
Minute volume
Total
MV
Mandatory
MVmand
Spontaneous
MVspon
Range
0 to 40 L/min
Accuracy for gas measurement with PGM
±10 % of the measured value or ±0.1 L/min
(the larger value applies)
Accuracy without PGM
±20 % of the measured value or ±0.2 L/min
(the larger value applies)
Resolution of displayed value
0.01 L/min (MV <1 L/min) or 0.1 L/min
(MV
≥
1 L/min)
16.13 Measuring
systems
and displays (continued)