Troubleshooting
!! 150
Emergency air inlet acti-
vated
There was not enough gas to
ventilate the patient. To main-
tain a minimum ventilation,
the device has started to use
ambient air.
Refill the breathing bag, e.g.,
with O
2
flush.
Increase fresh-gas flow.
Check the breathing circuit
for tight connections and
leakages.
An external anesthetic gas
monitor is extracting gas
from the breathing system
via a sample line.
Disconnect the sample line.
Seal the sample line port.
!! 135
etCO
2
high
etCO
2
has exceeded the
upper alarm limit.
Check ventilation.
!! 135
etCO
2
low
etCO
2
is below the lower
alarm limit.
Check ventilation.
!! 100
Exp. pressure sensor fail-
ure
Sensor calibration failed.
Ensure that a suitable substi-
tute monitoring is available.
Perform the system test.
If the problem persists, call
Dräger.
!
100
Expiratory flow sensor
expired
Accessory has been used
too long.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
acknowledge alarm.
!
255
Expiratory flow sensor
failure
The sensor is not calibrated.
The breathing system has
been replaced or discon-
nected since last calibration.
Perform the leakage test.
Failure of the flow sensor.
Replace the flow sensor and
perform a leakage test.
!!! 200
External fresh-gas outlet
failure?
The internal fresh-gas valve
is faulty. It is unclear whether
the fresh gas is being fed into
the internal breathing sys-
tem or towards the external
fresh-gas outlet.
Use "O
2
+" button to deter-
mine flow direction of fresh
gas: -If internal breathing
system or breathing bag fill,
external fresh-gas outlet is
not available. -If gas flows
out of the external fresh-gas
outlet, external fresh-gas out-
let can be used. Internal
breathing system can only be
used when the breathing bag
is not connected (ventilation
with ambient air only).
Check fresh-gas settings.
If the problem persists, call
Dräger.
Priority
Alarm
Cause
Remedy