Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
77
Assembly and preparation
Only use undamaged filters.
4. Attach the CO
2
absorber (8) to the breathing system from below. Rotate it in the
direction of the arrow until it reaches the stop.
Follow the instructions for use for the particular soda lime.
4.9.3 Water
trap
►
Empty or replace the water trap according to its instructions for use.
4.9.4
Connecting the sample line
►
WARNING
Risk due to occluded components in the breathing circuit
In the following cases, the sample gas flow can immediately cause negative
pressure in the lungs:
– Filters, hoses, or endotracheal tubes are blocked.
– The sample gas is being extracted between the patient and an occluded
component.
Ensure the following when ventilating pediatric patients and neonates:
►
When fine pored filters are used, do not connect the sample line between the
tube and the filter.
►
If the sample line is connected close to the patient, set the alarm limits for
MV low
and
Paw high
to suitable values. Observe the following information:
measurement by means of an HME filter", page 308.
WARNING
Risk due to incorrectly connected sample line
If the sample line is connected to the wrong connectors, e.g., connectors on
infusion pumps, fluids may be drawn in instead of sample gas. Consequently,
the gas measurement may not display correct values.
►
When connecting the sample line, take care that it is correctly connected.
WARNING
Risk of incorrect measured gas values
Blocked water traps or blocked sample lines prevent correct gas measurement.
As a result, incorrect measured gas values could be displayed.
►
Only use Dräger sample lines.