Chapter 4
:
Replacing Components and Preventive Maintenance
76
SE505 Hardware Manual
Clean the Air Regulator Filters
➢
To clean air regulator filters
1. Turn off all electrical power to the system. For more information, refer to the section
2. Open the front door of the machine.
3. Locate the air regulator inside SE505 at the lower right side of the machine.
4. Close the air shut-off valve to release the air pressure.
5. Unscrew bowl at the bottom of the regulator and take out the filter.
6. Rinse the filter in warm water and let it air dry.
7. Replace the filter and bowl at the bottom of the regulator.
Figure 56.
Air Regulator
Air shut-off valve
Air supply fitting
Bowl with filter
Summary of Contents for SE505
Page 1: ...SE505 Hardware Manual ...
Page 6: ...Contents vi SE505 Hardware Manual ...
Page 10: ...x SE505 Hardware Manual ...
Page 72: ...Chapter 3 Installation 62 SE505 Hardware Manual ...
Page 168: ...Chapter 4 Replacing Components and Preventive Maintenance 158 SE505 Hardware Manual ...
Page 180: ...Index 170 SE505 Hardware Manual ...