Wiring SE505 for Power
Wiring SE505 for Power
The SE505 system must be hard-wired into your AC main power. Have a certified electrician to
wire the system.
➢
To connect AC power
1. On the rear of the machine, turn off the AC main circuit breaker (CB1).
2. Loosen and remove four screws on the power access cover to remove the cover.
3. Loosen the strain-relief nut at the bottom of the box.
Note:
Make sure the strain-relief nut is fully tightened.
Figure 37.
Rear Electrical Panel
4. Feed the incoming power lines through the strain-relief connector.
Caution:
Before wiring the system, make sure that the AC main power supply is turned off and
locked off.
i
Note:
You must have a readily accessible power supply and a protective device in the fixed
wiring for the permanently-connected SE505 system. The power supply, wiring, and any
protective device must be rated at 240 VAC, 10 amps or 110 VAC, 15 amps.
Power access cover
(Loosen) Screws (x4)
(Loosen) strain relief nut
AC main
circuit
breaker
(CB1)