CyberOptics Corporation
43
Wiring SE505 for Power
Turning on System Power
CyberOptics recommends a 30–60 minute warm up time to ensure the most accurate system
measurements.
➢
To turn on the system
1. Check to make sure the conveyor is clear of panels and the E-Stop button is engaged.
2. Turn the AC main circuit breaker (CB1) to ON to turn on power to the entire system.
The breaker is located on the rear electronics panel.
3. Turn on the power switch on the front of the system.
4. Release the E-Stop button.
5. Press the green start button to restart.
Figure 39.
AC Main Circuit Breaker (CB1) Location
i
Note:
The computer will start automatically.
i
Note:
Do not run any Solder Paste Inspection applications until after the lights on the signal
light pole have turned on and then off. When you first start the SE505 system, the internal
processor initializes the system.
After approximately two minutes the lights turn on, and then several seconds later they turn off,
indicating that the initialization is complete. Also, there is a need to wait for the sensor light to
turn ON.
AC main circuit breaker (CB1)