When the temperature goes down and reaches the start temperature, the minimum on time is maintained
until the stop temperature is reached. Cool cells (or cooling pads) could operate according to outside tem-
perature. In such a case, the operation is exactly the same as described above, except that the tempera-
ture that EDGE considers is the outside temperature instead of the inside temperature,
Each cooling system operates primarily according to the average temperature of its respective group of
assigned inside sensors. An exception to this is that the set point of a cooling cell or pad system can be a
target outside temperature, where the outside temperature directly controls the operation of the system.
Some prerequisites or some optional conditions might have to be met for the cooling system to operate.
Amongst these conditions, there is animal age, outside temperature and time of day.
Current sensing
Current sensing can be used for the following reasons:
• To make sure the pump functions when it is set to function
• To make sure the pump doesn’t run dry
• To validate the operating time
• To make sure valves are activated or deactivated when they are set to activate or deactivate in order
to coordinate valves and pump operations. For example, pumps should not run if the water supply
valve is not powered on.
Cooling page and tile information
All the information concerning cooling equipment is displayed by pressing on a cooling tile in a house/
barn where cooling equipment is configured.
Cooling tile information
Cooling tile displays the number of cooling outputs currently running
Cooling page information
The following information is displayed in a cooling page
:
Cooling
device
Name of the cooling system (cool cell, fogger, high pressure mist, or mist) as entered when the
equipment was configured
Status
Status of the cooling device (bypassed, trouble, alarm, relative humidity compensation, on/off).
Note that only misting systems are used to increase relative humidity
Start
Temperature at which a cooling device starts at its minimum operating time
Stop
Temperature at which a cooling device stops (turns off). Stop temperature must be at least 0.5°
lower than start temperature
Max (in table
on left hand
side)
Temperature at which a cooling device reaches its maximum operating time.
T°
Average temperature currently measured by the assigned sensors
Run 24h
The amount of time the cooling device has been On in the past 24 hours
Control
probes
The names of the sensors used by the selected output, as entered during their configuration
144
895-00693
EDGE
Summary of Contents for EDGE 890-00601
Page 16: ...Chapter 1 Introduction 16 895 00693 EDGE...
Page 22: ...NOTES 22 895 00693 EDGE...
Page 26: ...Chapter 2 Basic connections Figure 2 1 EDGE 3 Slot Expansion Box 26 895 00693 EDGE...
Page 44: ...NOTES 44 895 00693 EDGE...
Page 52: ...NOTES 52 895 00693 EDGE...
Page 60: ...NOTES 60 895 00693 EDGE...
Page 76: ...NOTES 76 895 00693 EDGE...
Page 108: ...NOTES 108 895 00693 EDGE...
Page 110: ...Chapter 11 Test mode 110 895 00693 EDGE...
Page 128: ...NOTES 128 895 00693 EDGE...
Page 132: ...NOTES 132 895 00693 EDGE...
Page 138: ...NOTES 138 895 00693 EDGE...
Page 140: ...NOTES 140 895 00693 EDGE...
Page 142: ...NOTES 142 895 00693 EDGE...
Page 146: ...NOTES 146 895 00693 EDGE...
Page 150: ...NOTES 150 895 00693 EDGE...
Page 154: ...NOTES 154 895 00693 EDGE...
Page 156: ...NOTES 156 895 00693 EDGE...
Page 158: ...NOTES 158 895 00693 EDGE...
Page 176: ...Chapter 26 Test mode 176 895 00693 EDGE...
Page 188: ...NOTES 188 895 00693 EDGE...
Page 192: ...NOTES 192 895 00693 EDGE...
Page 194: ...NOTES 194 895 00693 EDGE...
Page 202: ...NOTES 202 895 00693 EDGE...
Page 204: ...NOTES 204 895 00693 EDGE...
Page 206: ...NOTES 206 895 00693 EDGE...
Page 214: ...NOTES 214 895 00693 EDGE...
Page 216: ...NOTES 216 895 00693 EDGE...
Page 218: ...NOTES 218 895 00693 EDGE...
Page 220: ...NOTES 220 895 00693 EDGE...
Page 224: ...NOTES 224 895 00693 EDGE...
Page 226: ...NOTES 226 895 00693 EDGE...
Page 230: ...NOTES 230 895 00693 EDGE...