Manual for VMP300-based instruments
97
Maximum Current
resolution
0.004 % of the range (down to 76 aA on the ± 1 pA)
Applied Current
Accuracy
< ±0.1% of range ±0.03% of setting for ±500 mA to ±100 nA ranges
< ±0.1% of range ±1% of setting for ±10 nA range to ±1 nA ranges
< ±0.2% of range ±2% of setting for ±100 pA range
Current Measurement
Ranges
± 100 pA, ± 1 nA, ± 10 nA, ± 100 nA
Additional
Ranges
with gain
± 1 pA, ± 10 pA
Maximum Resolution
0.004 % of the range
Accuracy
< ±0.1% of range ±0.03% of setting for ±500 mA to ±100 nA ranges
< ±0.1% of range ±1% of setting for ±10 nA range to ±1 nA ranges
< ±0.2% of range ±2% of setting for ±100 pA range
< ±1% of range ±2% of setting for ±10 pA range
< ±10% of range ±2% of setting for ±1 pA range
Impedance specifications (EIS option only)
Accuracy
10.2.3 1A/48V booster
Control amplifier
Compliance voltage
± 49V
Compliance current
± 1 A
Control voltage
± 48V
Bandwidth (-3dB)
>2MHz
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...