Manual for VMP300-based instruments
83
13) If all the items pass, the value of the calibration is saved in the memory of the board
and/or in the ULC electrometer. Click on
”OK” button to complete the calibration
process.
14) If one or more items failed, a warning
message indicating that the value of the
calibration will not be saved and the
calibration process is not completed. Before
contacting your local representative, please
check again if all the cables were properly
connected.
15)
If you selected “Channel” calibration the calibration procedure is finished. If not, the
calibration procedure continues with the calibration of the amplifier board. As this part
calibrates only the amplifier board, so the amplifier board has to be connected as well.
If user select “Amplifier” calibration”, the calibration procedure starts from this step.