Manual for VMP300-based instruments
37
4.4 Windows Security Alert
When the user tries to find
an instrument in the network
or by USB, the software will
use a broadcast that may be
stopped
by
windows
firewall. In this case click on
the “Allow access” button:
Fig. 44:
“Windows security Alert” window.
4.5 Firmware Upgrading with EC-Lab
software
When the user receives a new unit from the factory, the software and firmware (in the
instrument) are installed and upgraded. The instrument is ready for use. It does not need to be
upgraded. However, when a new EC-Lab
version is released (with new protocols or
improvements) the firmware has to be updated and installed by the user.
4.6 Firmware Downgrading with EC-Lab
software
It is possible to downgrade the firmware of the instrument in an advanced tool available in EC-
Lab
. This procedure should only be done by an advanced user and in very special cases.
Please contact technical support for assistance.
WARNING:
If the user downgrades the firmware of the instrument, corresponding version of EC-
Lab
must be used. For example, if the firmware of the instrument is the V10.20, the
user must control the instrument with EC-Lab
V10.20.
Make sure that the older version supports the particular instrument.
Instruments
Lowest compatible version
SP-200
V10.00
SP-240
V10.17
SP-300
V10.00
VSP-300
V10.17
VMP-300
V10.30
The firmware downgrade procedure is as follows:
NOTE:
the example shows a downgrade from V10.23 to V10.12 but the procedure will be the
same for other versions.
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...