15
2.1.3.2 HCV-3048
WARNING
In HCV-3048, the air flows from front to the rear of the instrument (Fig. 13-14).
It should be
placed at a minimum of 30 cm from a wall.
Otherwise, the lifetime and the specifications of
the instrument are not guaranteed.
Fig. 11: Air flow for HCV-3048
WARNING
As a general rule, the instruments should not be seated on an inflammable surface.
Please prefer inert surface such as glass or stone. This does not apply for instruments
connected in parallel, where they would be vertically stacked.
2.2 Unpacking and inspection
. VSP-300, VMP-300 and HCV-3048 are heavy instruments (22, 30
and 26 kg, respectively) and necessitate two persons to unpack and
carry them, to avoid crashing and health hazard.
2.3 Electrical equipment requirements
Please find below the electrical specifications of the instruments:
Table 5: Electrical input specifications of the instruments described in this manual.
Instrument
Input
Voltage Range
/Vac
Frequency
range/Hz
Power max
/W
Fuses
a
SP-200
[90;264]
[50;60]
350
2 x 4 AF
SP-240
[90;264]
[50;60]
350
2 x 4 AF
SP-300
[90;264]
[50;60]
350
2 x 4 AF
VSP-300
[90;264]
[50;60]
650
2 x 5 AF
VSP-300
(Vertical)
[90;264]
[50;60]
650
2 x 5 AF
VMP-300
[90;264]
[50; 60]
1500
2 x 10 AF
HCV-3048
[85;264]
[47;40]
2000
2 x 16 AF
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...