48
5.2.6 Floating mode
The instrument’s floating mode is selected in
the advanced settings window of the EC-Lab
®
software. Both the standard mode and the
floating mode can be used without modifying
the cell connection.
Fig. 67: Floating or Grounded mode
selection.
The floating mode allows the user to work with grounded cells, autoclaves, pipelines, glove
boxes, etc.
When the channel board is in floating mode, the red LED is not bright as it is when the channel
board is in Grounded mode.
Note that some specifications of the instrument can be modified using the floating mode.
Using the floating mode for measurements will require taking some measurement precautions.
Indeed, the cell connections should be isolated from the ground such as other apparatus that
can be connected to the cell (multi-
tester, voltmeter …).
5.3 Specific connection for high current boosters.
5.3.1 Connection to the cell with 1A/48V booster
The connection is similar to the connection described in the previous paragraph. The only
difference is that the potentiostat/galvanostat (with/without EIS) board has to be connected to
the 1A/48 V booster with the DB15 connection cable.
Fig. 68: Channel board connection to the cell with the 1A/48V booster.
P1 + S1
P2 + S3
S2
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...