Manual for VMP300-based instruments
40
8) The downgrade
process
starts
and
“Upgrading...”
appears on the
bottom of the
window (but the
firmware is really
downgraded).
Fig. 52: Flash loading.
Fig. 53: Upgrading message on the status bar.
9)
Close the “firmware downgrading/upgrading” window and EC-Lab
V10.23
. The
downgrading procedure is now completed and you can connect the older version to the
instrument.
10) Open the older version of EC-Lab
V10.12
and connect to the instrument.