Manual for VMP300-based instruments
75
8.1.1 Channel board without EIS option
The procedure is the following:
- Load the CV technique with the default
settings on the channel of interest. The
screenshot of the settings is shown in Fig. 108.
NOTE:
For longer cable the bandwidth parameters
may be reduced [3].
-
Start the measurement.
-
Select the whole resulting curve
-
Fit the resulting data with the “
Linear
Fit
” tool available in the “
Analysis
”
menu (or by clicking on “Shift+F2”).
Fig. 108: Default settings of the CV
technique.
TIP:
To get directly the value in Ohm
([Volt]/[Amperes]), you should shift the I
vs.
E
we
representation
(default
representation for CV) to E
we
vs.
I
representation. This shift can be easily
done thanks to the representations tool
available at the top of the Graphic frame
(Fig. 109).
Fig. 109: Ewe vs. I representation.