Manual for VMP300-based instruments
14
Fig. 9: Air flow for SP-200/SP-240/SP-300/Horizontal VSP-300 and VMP-300 chassis.
WARNING
There should be sufficient space at the rear of the instrument to prevent the hot air from
reentering the instrument.
A minimum distance of 30 cm should be observed.
It must also be ensured that nothing can block the air flow at the bottom of the instrument.
Something as harmless as a sheet of paper can reveal itself quite dangerous if it is stuck
underneath the instrument, on the ventilation grids, blocking the air flow.
For the vertical VSP-300 it must be ensured that sufficient space if left on both sides of the
instrument.
Again a minimum distance of 30 cm should be observed.
Fig. 10: Air flow for vertical VSP-300