Manual for VMP300-based instruments
58
Table 7: Help for bandwidth selection for a standard board with a booster considering
the capacity and the time constant (TC) of the studied cell, but also the experimental
conditions. Considering these factors the most appropriate bandwidth (SF) is given.
1 µA/
1 MΩ
10 µA/
100 KΩ
100 µA/
10 KΩ
1 mA/
1 KΩ
10 mA/
100 Ω
100 mA/
10 Ω
1 A/
1 Ω
TC
SF
TC
SF
TC
SF
TC
SF
TC
SF
TC
SF
TC
SF
10 pF
10 µs 7/8
1 µs
1-9 100 ns 1-9 10 ns 1-9
1 ns
1-9 100 ps 1-9 10 ps 1-9
100 pF
100 µs 7
10 µs 7/8
1 µs
1-9 100 ns 1-9 10 ns 1-9
1 ns
1-9 100 ps 1-9
1 nF
1 ms
6 100 µs 7
10 µs 7/8
1 µs
1-9 100 ns 1-9 10 ns 1-9
1 ns
1-9
10 nF
10 ms
5
1 ms
6 100 µs 7
10 µs 7/8
1 µs
1-9 100 ns 1-9 10 ns 1-9
100 nF
100 ms 4
10 ms
5
1 ms
6 100 µs 7
10 µs 7/8
1 µs
1-9 100 ns 1-9
1 µF
1 s
3 100 ms 4
10 ms
5
1 ms
6 100 µs 7
10 µs 7/8
1 µs
1-9
10 µF
10 s
2
1 s
3 100 ms 4
10 ms
5
1 ms
6 100 µs 7
10 µs 7/8
100 µF
100 s
1
10 s
2
1 s
3 100 ms 4
10 ms
5
1 ms
6 100 µs 7
1 mF
1
ks
100 s
1
10 s
2
1 s
3 100 ms 4
10 ms
5
1 ms
6
10 mF
10
ks
1
ks
100
1
10 s
2
1 s
3 100 ms 4
10 ms
5
100 mF
100
ks
10
ks
1
ks
100 s
1
10 s
2
1 s
3 100 ms 4
1 F
1
Ms
100
ks
10
ks
1
ks
100 s
1
10 s
2
1 s
3
10 F
10
Ms
1
Ms
100
ks
10
ks
1
ks
100 s
1
10 s
2
100 F
100
Ms
10
Ms
1
Ms
100
ks
10
ks
1
ks
100 s
1
6.2 Floating mode
By default, the instrument is configured in
grounded mode. Changing this configuration
can be done in the Advanced Settings window
of EC-Lab
®
software. Both the standard and
floating mode can be used without modifying
the cell’s connection.
Fig.
81:
Floating/Grounded
mode
selection frame.
Note: In EC-Lab, when the channel board is in floating mode, the red LED is not bright as it is
when the channel board is in Grounded mode.
6.3 Filtering
In some experimental conditions, it would
be useful to use a filter, then three
possibilities of current and potential filter
are available with the instrument. The filter
configuration can be done in the Advanced
settings window of EC-Lab
®
software.
Note the two gains are only available if the
one of the filter is activated.
Fig. 82: filters selection frame.
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...