Manual for VMP300-based instruments
42
Note that when the channel board is running, the green LED blinks.
5.1.2.2 Booster boards
There are two kinds of booster boards. The booster boards which ne
edn’t any additional cable
to carry the current (1A/48V) and the boosters which need a cable to carry the current (2A/30V;
4A/14V; 10A/5V). The 2A/30V; 4A/14V; 10A/5V boosters can be connected in parallel to
increase the maximum current.
NOTE:
It is possible to connect in parallel only the same boosters, not possible to mix 4A and
10 A booster for example.
Fig. 55: Booster boards.
5.1.2.2.1 1A/48V booster board
DB-15 female connector for connection to the calibration board,
DB-15 male connector for connection to the potentiostat board.
The 1A/48V booster increases the unit’s compliance and the cell polarization to +/- 48 V.
This option is compatible with the Ultra-Low current option. This combination of options allows
user to get a current autoranging from 1 pA to 1 A. This option can be used both with EIS
(maximum value 2 MHz) and standard channel boards. Once the booster is connected to the
potentiostat board, connection to the cell is the same as without the booster 1A/48V connected.
5.1.2.2.2 2A/30V; 4A/14V; 10A/5V booster board
DB-15 female connector for connection to the calibration board,
DB-15 male connector for connecting the potentiostat/galvanostat board to the booster
board,
DB-9 connector for connecting the power cable P1 and P2 leads.
The 2A/30V, 4A/14V and 10A/5V boosters have similar boards with one DB-15 female
connector, one DB-15 male connector and one DB-9 male connector. These boosters are
connected to channel board and to calibration board in the same way.
The booster connection is as follows:
Connect it to the standard channel with the DB15 male
– DB15 female cable
The power cable with P1 and P2 leads of the kit has to be connected to the booster
board
via
DB9 connector.
Note that the connection to the cell is the following:
P1 lead of the standard or ultra-low current (ULC) board cable has to be connected to
the P1 lead of the booster option.
P2 lead of the standard or ultra-low current (ULC) board cable has to be connected to
the P2 lead of the booster option.
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...