Manual for VMP300-based instruments
94
10.
Technical Specifications
10.1 Equipment Ratings
10.1.1 Electrical & Mechanical specifications
Table 10: Instrument electrical ratings.
Instrument
type
Input
Output
Mechanical
Voltage
Range/
Vac
Frequen
cy
range/H
z
Power
max/
W
Fuses
a
Voltage Range/Current
Size
(HxWxD)/mm
Weight
without
cable/kg
SP-200
[90;264]
[50; 60]
350
2 x 4 AF
10 Vdc / 500 mA
221x170x384
7.2
SP-240
[90;264]
[50; 60]
350
2 x 4 AF
10 Vdc / 500 mA
or 14 Vdc / 4A (with 4A/14V
booster)
225x205x410
7.5
SP-300
[90;264]
[50; 60]
350
2 x 4 AF
10 Vdc / 500 mA
or 48 Vdc/1A (with 1A/48V booster)
or 30 Vdc/2A (with 2A/30V booster)
or 14 Vdc/4A (with 4A/14V booster)
or
5
Vdc/10A
(with
10A/5V
booster)
225x205x410
7.5
VSP-300
[90;264]
[50; 60]
650
2 x 5 AF
260x530x510
22
VSP-300
(Vertical)
[90;264]
[50; 60]
650
2 x 5 AF
337x254x517
20
VMP-300
[90;264]
[50; 60]
1500
2 x
10 AF
533x315x563
30
HCV-3048
[85-264]
[47-440]
[1700-
2000]*
2 x
16
AF
[0;48] Vdc / ±30 A
135x400x430
26
* See power contour plot of HCV-3048
10.1.2 Environmental
-
Indoor use
-
Operating Temperature:
10°C to +40°C Indoor use
-
Storage Temperature:
0°C to +50°C
-
Pollution degree:
1 (no pollution or only dry)
-
Altitude:
<2000 m above sea level
-
Humidity:
10% to 80% non-condensing
-
Case protection:
IP20
-
Warm-up:
1 hour to rated accuracy
-
Cooling:
Internal DC Fans
-
Vibration:
not specified
-
Choke:
not specified
-
Noise (audible) :
<60 dBa at full power
Safety
complies with EN61010-1.
EMC
complies with EN61326.
RoHS
complies with 2002/95/EC
10.2 Channel specifications
10.2.1 Channel board
General Functions
Potentiostat
Yes
Galvanostat
Yes
Impedance analyzer
Yes (option)
Coulometer
Yes
Linear scan generator
Yes (option)
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...