79
8.2 Channel calibration with EC-Lab
®
software
8.2.1 Potentiostat/galvanostat board calibration
To perform the calibration the user has to follow the procedure described in detail hereafter.
The calibration procedure lasts a few minutes.
1) In EC-Lab
software
Select "
Tools/channel
calibration
". You will
receive the following
message.
2) Answer "
Yes
" to the
warning
message.
Then
the
board
calibration
window
appears.
3) Select the channel you want to calibrate with the drop-down box
4) Select the type of the calibration you want to run with the drop-down box: three different
calibrations are offered
i.e.
only the
“Channel”, the “Channel & Amplifier”, or only the
“Amplifier”
.
If you want to calibrate:
-
the channel board alone, select “Channel”
- the booster and the channel
board, select “Channel & Amplifier”
-
the booster alone, select “Amplifier”
5)
Click on the “Calibrate” button and this warning message appears which indicates that
if you start the calibration all the settings of the channel selected will be reset.
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...