Manual for VMP300-based instruments
47
5.2.3 CE to Ground connection mode
This connection mode is chosen in the
software “Advanced settings” window. Then
the connections must be done in a special
way, in connecting the ground cable to the
CE electrode of the cell, as shown below:
Fig. 64: CE to Ground connection mode.
5.2.4 WE to Ground connection mode
This connection mode is chosen in the
software “Advanced settings” window. Then
the connections must be done in a special
way, in connecting the S1 and the Ground
cable to the WE of the cell and S3 and P1 to
the CE electrode of the cell, as shown below:
Fig. 65: CE to Ground connection mode.
5.2.5 High voltage control connection mode
This connection mode is chosen in the software “Advanced settings” window. Then the
connections must be done in a special way, in connecting the S2 and the P3 cable to the WE
of the cell and S1 and P1 to the CE electrode of the cell, S2 is connected to RE. Connection
is shown below:
NOTE: when this mode is selected the impedance techniques are no more available.
Fig. 66: High voltage control connection mode. Left: +/- 48V for the 1A/48V booster.
Right: +/- 30V for the 2A/30V booster.
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...