Manual for VMP300-based instruments
45
E
we
=
S1
-
S2
E
ce
=
S3
-
S2
Notes
:
CAUTION
Disconnection of the cell cable without switching off the instrument can damage the
electrometer located on the cable (either on the standard cable or on the ultra-low
current option)
WARNING
Voltage limits for standard and ULC cables:
-
In Grounded mode: the potential between the grounded lead (black lead) and the
terminals should not exceed 50 V. The potential between the grounded lead and the
earth not exceed 10 V.
-
In Floating mode, the potential between the ground lead (black lead) and the terminals
should not exceed 50 V. The potential between the ground lead and the earth should
not exceed 100 V.
Fig. 59: voltage limits in grounded and floating mode.
The current (defined in the positive direction) crosses the electrochemical cell from P1 to P2.
Three typical standard configurations are explained below.
5.2.2 Standard connection
5.2.2.1 Standard three-electrode connection
In the standard three-electrode connection mode
typically used in analytical electrochemistry or
corrosion experiments, the working electrode is
connected to S1+P1. The counter-electrode is
connected to S3+P2, and the reference electrode
is connected to S2.
Fig. 60: Standard three-electrode
connection for a classical metal-
solution interface.
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...