Manual for VMP300-based instruments
55
Trigger In
and
Trigger Out
are techniques
programmed into the experiment protocol. . Trigger
In can be used to start or stop an electrochemistry
technique. For this, another instrument must send
a trigger to the instrument. Trigger Out can be sent
at the beginning or the end of the experiment. It is
used to start or stop an experiment on another
instrument.
The Trigger signals have high and low levels:
+ 3.5 V < Trigger high level < + 5 V
0 V < Trigger low level < 0.8 V
Fig. 79: Structure of the DB9 connector.
Analog Out
is used to control an external device (1 mA max).
Analog In1
and
Analog In2
are general analog inputs used for auxiliary signal
(voltage) recording. Activate Record Aux1 and Record Aux2 in the cell characteristics
window to record these variables in the data file.
±10 V max, accuracy: 16 bits resolution with automatic ± 2.5 V, ± 5 V, ± 10 V ranges
for Analog In1, ±10V for Analog In2, Input impedance: 1 MΩ/GND + 10 kΩ in series.
E monitor
is a voltage analog output that can be used to inspect E
we
with an
oscilloscope. The output signal is reversed with respect to the software measured
values.
I monitor
is an voltage analog output that can be used to inspect I with an oscilloscope.
The current to voltage conversion depends on the Irange selected:
Current Range
S/W Current Reading
Std Cable
I mon Output
ULC Cable
I mon Output
1A
±500mA (max)
±0.5V
±0.5V
100mA
±100mA
±1V
±1V
10mA
±10mA
±1V
±1V
1mA
±1mA
±1V
±1V
100uA
±100uA
±1V
±0.9V
10uA
±10uA
±1V
±1V
1uA
±1uA
±1V
±0.9V
100nA
±100nA
±0.1V
±0.9V
10nA
±10nA
±0.01V
±0.9V
1nA
±1nA
x
±0.9V
100pA
±100pA
x
±1V
10pA
±10pA
x
±0.1V
1pA
±1pA
x
±0.01V
OPEN In
is an external Trigger signal (active low) that opens relays to Open Circuit
and stops the experiment performed on the cell. This may be used as an emergency
stop of the experiment that is triggered by an external monitor/event.
Ground
is tied to the earth.
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...