Manual for VMP300-based instruments
101
Accuracy
10.2.7 HCV-3048 used with HV-48 electrometer
Control amplifier
Compliance voltage
[-1,50] V
Compliance current
± 30 A
Current ranges
3 (30 A, 3 A, 300 mA)
Control voltage
[0,48] V
Accuracy
User (using calibration board,
when HCV-3048 purchased
separately from pstat board)
Voltage control
0.1 % + 0.1 % (range + reading)
Current control
0.3 % + 0.5 % (range + reading)
Voltage measurement
0.1 % + 0.1 % (range + reading)
Current measurement
0.3 % + 0.5 % (range + reading)
Factory (when HCV-3048
purchased with pstat board, or
pstat board returned to factory
for calibration with HCV-3048)
Voltage control
0.03 % + 0.03 % (range + reading)
Current control
0.1 % + 0.3 % (range + reading)
Voltage measurement
0.03 % + 0.03 % (range + reading)
Current measurement
0.1 % + 0.3 % (range + reading)
Bandwidth (-3dB)
120 kHz
Operating area
Control/measurement : [-30,30] A @ [0, 48] V
(with 2.5 m cables)
Slew rate (no load)
20 V/µs
Rise/Fall time (no load)
3 µs
Floating mode
Yes
(350 kΩ isolation)
Parallel ability
Yes up to 4 (120 A)
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...