Manual for VMP300-based instruments
96
Input impedance
1
TΩ ΙΙ 25 pF typical
Input Bias current
< 10 pA
Bandwidth (-3 dB)
8 MHz
Common mode rejection ratio
> 60 dB at 100 kHz
Ground to chassis impedance
Floating mode
1
MΩ ΙΙ 25 pF typical
Grounded mode
< 100
Ω
IR compensation
Resistance determination
EIS
Compensation mode
Hardware or software positive feedback
Compensation range
Programmable from 0 to 100 % of the current range
resistor
Auxiliary Inputs/outputs
External Input
Can be used to apply an external waveform directly to
the control amplifier
2 Analog Inputs
Automatic ± 2.5 V, ± 5 V, ± 10 V ranges - 16 bits
resolution
1 Analog Output
16 bits resolution on the ± 10 V range
2 Digital Inputs
TTL level Trigger input and Open input
1 Digital Output
TTL level Trigger output
2 Monitor Outputs
Cell current and compensated working electrode
potential
Impedance specifications (EIS option only)
Frequency range
10 µHz to 7 MHz
Frequency resolution
< 10 ppm of the setting
Sinus amplitude
0.5 mV to 5 V with 76 µV resolution
0.1 % to 100 % of the current range with resolution of
0.004 % of the range
Accuracy
See contour plot
Mode
Single sine, Multisine, FFT analysis
Contour plot
10.2.2 Ultra Low Current
Cell Control
Maximum Current
± 500 mA (± 1 A continuous with 1 A/48 V)
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...