Manual for VMP300-based instruments
13
2.
Getting started
2.1 Hardware installation
2.1.1 Preparing the physical installation
Questions
YES NO
Has the incoming electrical power been taken into account ? (1)
Have the environmental specifications been taken into account ? (Room
temperature, room humidity, pollution degree, EMC environment...) (1)
Have the electrical connections cables been supplied?
(1): See 10.1 Equipment Ratings
If all the answers are YES then you can proceed with the installation.
2.1.2 Safety emergency shutdown
As described in Part 1.1.2, HCV-3048 is a high power instrument and features an
Emergency
Stop
button. This button will
This button will open the output of the power at the power
sockets
.
WARNING
When HCV-3048 are used in parallel to increase the delivered current, it is possible to
simultaneously open the relay of the power output on all the boosters parallelized with only
one
Emergency Stop
by connecting the BNCs in the back of all four instruments.
The connection can be made by using BNC T-connector on the Open In of all boosters except
the main one on which the potentiostat is connected (See 5.1.3 for more information on
connections).
2.1.3 Ventilation requirement
As our instruments are delivering power, they produce heat which needs to be evacuated to
maintain the temperature inside of the instrument as constant as possible. Precautions must
be taken to avoid blocking the air flow. Blocking the air flow will increase the temperature of
the instrument, hence reducing the lifetime of the components as well as impairing its
capabilities. All the specifications and capabilities listed in this manual are guaranteed under
recommended operation, which means normal air flow
2.1.3.1 Potentiostats/galvanostats
In all our instruments (except the vertical VSP-300) the air flows from the bottom or the front
to the back of the instrument (Fig. 9).
For the vertical VSP-300 the air flows from the right to the left when facing the instrument
(Fig. 10).
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...