Manual for VMP300-based instruments
66
7.
Options & Accessories
The options and accessories of the VMP300-based instruments are described in this section.
7.1 Analog Ramp Generator (ARG)
This option is an electronic board which
has
to
be
plugged
onto
the
potentiostat/galvanostat board. It allows
users to apply a true linear voltage scan
(analog and not digital) to the cell.
The analog nature of the scan is critical in
several applications such as catalyst study
for fuel cell development or for voltammetry
requiring fast scan rate (to detect
intermediate species with short lifetime).
As the linear scan generator allows fast
scan investigation, hardware ohmic drop
compensation is also implemented in the
software.
This
correction
is
done
continuously
by
the hardware which is
much
faster
than
the
software
compensation. For example, we can
compare the standard and linear CV. Due
to the feedback loop, the software
compensation with the standard CV takes
around
100
µs
while
the hardware
compensation is about 100 times faster
.
This option is compatible with ULC cable
and booster boards.
Fig. 93: ARG option.
ARG specifications:
Scan ranges
1 MV/s, 10 kV/s, 100 V/s, 1 V/s.
Scan resolution
0.0015% FSR* (down to 15 μV/s).
Minimum voltage
scan rate
0.1 mV/s
Sampling rate
down to 1 µs
Buffer size
for Fast mode
40 000 data points
Voltage range
± 10 V. The voltage range can be
adjusted.
Accuracy
< ±0.1% of range
Number of cycles
1 to 65535
7.2 Isolation System (IS1)
It is used to keep a potentiostat totally floating even when it is connected to a grounded external
device via the auxiliary cable.
The auxiliary cable is connected to the potentiostat board using the DB9 (or SubD 9)
connector. Thanks to this cable, an external device can communicate with the potentiostat
either by sending/receiving digital triggers or sending/receiving analog signals. When the
auxiliary cable is connected to an external device, the potentiostat board becomes grounded
to the internal ground of the external device. Consequently, if the channel was initially used in
the floating mode, connecting it to the external device using the auxiliary cable makes it lose
its floating character.
Summary of Contents for HCV-3048
Page 2: ......
Page 97: ...Manual for VMP300 based instruments 89...
Page 98: ...Manual for VMP300 based instruments 90 The next steps are similar to the in built boosters...
Page 114: ...Manual for VMP300 based instruments 106 12 EC declaration of conformity 12 1 SP 200...
Page 115: ...Manual for VMP300 based instruments 107 12 2 SP 240 SP 300...
Page 116: ...Manual for VMP300 based instruments 108 12 3 VSP 300 12 3 1 Horizontal VSP 300...
Page 117: ...Manual for VMP300 based instruments 109 12 3 2 Vertical VSP 300...
Page 118: ...Manual for VMP300 based instruments 110 12 4 VMP 300...
Page 119: ...Manual for VMP300 based instruments 111 12 5 HCV 3048...