5 Alarms and other messages
5.3 Alarm inhibition
Pressing the alarm inhibition key interrupts the alarms for 2 minutes.
Press this key to cancel alarm inhibition. The alarm message remains on the screen.
However, if another audible alarm with a higher priority replaces it, the sound inhibition is automatical-
ly disabled and the audible emission of the new alarm begins. The previous alarm is then automatical-
ly archived in the history.
5.4 Preventive alarm inhibition
Holding down the alarm inhibition key interrupts the alarms for 2 minutes.
On the screen, the panel shows “Alarm inhibited”. A counter next to the alarm inhibition key shows the
remaining activation time.
If there is an alarm during this time, the alarm message displays on the screen.
Pressing this key a second time cancels the preventive inhibition.
The inhibited alarms concern the alarms relating to monitoring and patient disconnection.
In the event of a technical alarm (mains disconnection, etc.) or Pmax alarm, inhibition is disabled.
5.5 Reset
When an alarm threshold is exceeded, it continues to be displayed in red, even when the alarm disap-
pears.
Press the threshold displayed in red to return to the normal display.
5.6 History
This function displays the chronological list of the last 200 alarms or events recorded by the ventilator.
To access the history, see Patient monitoring on page 67.
For each alarm there is a date, priority (colour), time, activation or deactivation (ON or OFF), name, and
for physiological alarms, the alarm threshold setting at the time of activation.
Use the control wheel to browse the list by turning it clockwise or anti-clockwise.
To exit the alarm history, press the control wheel.
81