1 Before use
•
The replacing of cables or internal components with cables or components that are not supplied
by
Air Liquide Medical Systems
may result in an increase in emissions or a decrease in the immuni-
ty of the device.
Connection to other electrical devices
•
Do not connect the device to other electrical appliances not mentioned in this user manual without
first consulting the manufacturers concerned or a specialist.
•
Devices connected to the inputs and signal outputs must comply with the 60601-1 Standard.
Set-up
•
The device must not be put into service immediately after storage or transportation where the tem-
perature and humidity were different from the recommended operating conditions.
•
Before each use, check that the audible and visual alarms are working correctly and carry out the
checks listed in the appendix (
see section
“Checklist on page 139).
•
The ventilator should not be covered or positioned in such a way that its functioning or perfor-
mance are affected. Always leave some space around the device: for example, never place the
ventilator close to a curtain which could impede the fresh air flow and cause overheating.
•
If the
Monnal T60 Advanced
is installed on the universal support (KA010400), follow the instruc-
tions in the assembly manual. Ensure that the structure or unit (bed rail, for example) onto which
the universal support is mounted can support the weight of the
Monnal T60 Advanced
.
Use
•
The manufacturer has tried to anticipate most of the possible malfunction scenarios of this ven-
tilator, and these are normally monitored by the internal monitoring system. It is nevertheless rec-
ommended, in cases of complete patient dependence, that you provide an additional, fully au-
tonomous system which can be used to check the effectiveness of the ventilation, as well as a
back-up device, such as a suitable manual insufflator.
•
Lack of an alternative means of ventilation may result in patient death should the ventilator fail.
•
If the accessories used are not compliant with the manufacturer‘s recommendations, the manu-
facturer accepts no responsibility in the event of an incident.
•
Do not expose the device to direct sunlight.
•
Do not use
Monnal T60 Advanced
in a hyperbaric chamber.
•
The device and its accessories (masks, circuits, etc.) are Latex-free.
•
The air inlets located at the back and side of the device must be completely unobstructed.
•
To operate the device from ambient air, a Monnal Clean-In (HEPA) filter must be used at the venti-
lator inlet. This filter is recommended by
Air Liquide Medical Systems
.
•
Do not use the ventilator in an explosive or nicotine-laden atmosphere (cigarette smoke, fire, etc.).
•
In order to prevent dust from entering:
•
between ventilator uses in the bag, close the inspiratory limb cap;
•
between ventilator uses in the bag, leave a bacteriological filter or patient circuit on the inspira-
tory outlet of the ventilator;
•
during cleaning, leave a bacteriological filter or a patient circuit on the inspiratory outlet of the
ventilator;
•
clean the inside of the bag regularly.
Transportation
•
During transportation, we recommend that you use the device in its protective carry bag. The case
must be securely fastened in the vehicle using the strap loops provided for the purpose.
•
The device must not be subjected to violent impact.
•
Use the carry bag recommended by
Air Liquide Medical Systems
only.
13