Monnal T60 Advanced - User manual
3.4 Electrical power supply
Connect the electrical power supply cable to the ventilator, and then connect it to an AC power supply
socket.
Location of power cable with anti-pull cable protection clip :
CAUTION:
Always check that the electrical network is compatible with the specifications in
this manual.
Check that the power cable is not damaged.
CAUTION:
Leave enough space behind the device for removing the cord from the power out-
let when stopping the machine.
Note:
The power cord enables
Monnal T60 Advanced
to be disconnected from the power
supply on all poles simultaneously.
3.5 Oxygen supply
To use a mixture with more than 21% oxygen, connect the ventilator’s high- or low-pressure O
2
inlet to
a valid source via an appropriate connector.
If this oxygen source is a high pressure transport cylinder, it must be equipped with a pressure reducer
to suit the allowable pressure range (2.8 to 6 bar).
CAUTION:
Start by connecting the O
2
connection hose to the ventilator before connecting it
to the oxygen network.
CAUTION:
Check the capacity of the oxygen cylinder before using the ventilator.
20