3 Installation and commissioning
2.
Apply power to the ventilator.
3.
Connect the probe to its patient adapter (a). The probe is correctly connected to its adapter
when you hear it click into position.
4.
Wait at least 10 seconds. If the etCO2 monitoring block does not show 0%, or if the error ‘CO2
measurement incorrect’ is displayed, a calibration test should be carried out. See Display of
5.
The LED flashes and then goes green. This means that the IRMA
™
probe is ready for use (b).
Pre-use checks (for each new patient)
1.
Connect the sensor to the patient adapter (a). A click should be heard when the sensor is
properly connected to the adapter.
2.
Ensure that the etCO2 monitoring block is displaying data.
3.
Connect the IRMA
™
probe, equipped with its adapter, to the Y-piece on the patient circuit (c).
4.
Connect the IRMA
™
probe to the patient’s endotracheal tube (d).
5.
Position the IRMA
™
probe (e) (see photo opposite).
Note:
The probe sends information and alarms to the
Monnal T60 Advanced
. The probe has
an LED indicating the following statuses:
LED status
Description
Continuous green
OK
Flashing green
Calibration in progress
Continuous red
Probe error
Flashing red
Checking the adapter
23