![Air Liquide MONNAL T60 ADVANCED User Manual Download Page 134](http://html1.mh-extra.com/html/air-liquide/monnal-t60-advanced/monnal-t60-advanced_user-manual_2875003134.webp)
Monnal T60 Advanced - User manual
Adult
Child
Infant
Thresh-
old (name)
Ventilation modes
Set-
ting
by
Min~Max [default]
ECLS activated
max value (50 or
low VTe thr
50) ~ 3,000 [2,000]
max value (50 or
low VTe thr
50) ~ 2,000 [1,000]
NA
Pressure-con-
trolled modes:
(A)PCV, PSV, CPAP,
APRV, PSIMV, PRVC,
PS-Pro
OFF then 10 ~ min val-
ue (2,000 or high VTi
threshold - 50) [OFF]
OFF then 10 ~ min val-
ue (1,950 or high VTi
threshold - 50) [OFF]
OFF then 10 ~ min
value (790 or high VTi
threshold - 10) [OFF]
Volumetric modes:
(A)VCV, VSIMV
OFF then 10 ~ min
value (2,000 or
Vt setting) [OFF]
OFF then 10 ~ min
value (1,950 or
Vt setting) [OFF]
OFF then 10 ~
min value (790 or
Vt setting) [OFF]
Low inhaled
volume (VTi)
ECLS activated
10
OFF then 10 ~ min val-
ue (2,000 or high VTi
threshold - 50) [OFF]
OFF then 10 ~ min val-
ue (1,950 or high VTi
threshold - 50) [OFF]
Not available
Pressure-con-
trolled modes:
(A)PCV, PSV, CPAP,
APRV, PSIMV, PRVC,
PS-Pro
max value (50 or
low VTi thr
50) ~ 3,000 [2,000]
max value (50 or
low VTi thr
50) ~ 2,000 [1,000]
max value (20 or
low VTi threshold
+ 10) ~ 800 [100]
High inhaled
volume (VTi)
Volumetric modes:
(A)VCV, VSIMV
10
max value (50 or
Vt s 50)
~ 3,000 [2,000]
max value (50 or
Vt s 50)
~ 2,000 [1,000]
max value (20
or Vt s
10) ~ 800 [100]
All except Hflow
OFF then 0.1~39 [3]
OFF then 0.1~39 [1.5]
OFF then 0.1~39 [1.0]
Low exhaled
minute vol-
ume (MVe)
ECLS activated
0.1
OFF then 0.1~39 [OFF]
Not available
All except Hflow
1.5~40 [25]
1~40 [10]
1~40 [5]
High exhaled
minute vol-
ume (MVe)
CPV
0.1
1.5~40 [40]
Not available
All except Hflow
0.5~39 [3]
0.5~39 [1.5]
0.5~39 [1.0]
Low inhaled
minute vol-
ume (MVi)
ECLS activated
0.1
0.5~39 [3]
0.5~39 [3]
Not available
All except Hflow
1.5~40 [25]
1~40 [10]
1~40 [5]
High inhaled
minute vol-
ume (MVi)
CPV
0.1
1.5~40 [40]
Not available
Low FiO2 (FiO2)
All
18~95 [FiO2 setting -5]
High FiO2
(FiO2)
All
1
24~105 [FiO2 s5]
All except Hflow
OFF then 1~98 [30]
CPV
OFF then 1~98 [OFF]
Not available
Low etCO2
(etCO2)
ECLS activated
1
OFF then 1~98 [OFF]
Not available
High etCO2
(etCO2)
All except Hflow
1
5~99 [49]
High Pplat
(Pplat)
All except Hflow
1
1~50, OFF [OFF]
Low Chest
Compression
Frequency (fCC)
CPV
5
70~100 [90]
Not available
134