Monnal T60 Advanced - User manual
Other modes
When switching from one ventilation mode to another (excluding emergency modes and excluding
ECLS), the values of the settings and of the thresholds are saved. For the settings and thresholds that
are not common to the two modes (for example, the VT in a pressure-controlled mode), the values are
memorized but not applied.
In this example, a
pressure
-controlled mode (PSV) follows a
volume
-controlled mode ((A)VCV):
(A)VCV
→
PSV
→
PSV
→
(A)VCV
VT: 460 mL
PEEP: 4 cmH
2
O
PEEP: 4 cmH
2
O
PEEP: 6 cmH
2
O
VT: 460 mL
PEEP: 4 cmH
2
O
1.
Ventilation in
(A)VCV
mode with a VT set at 460 mL and PEEP of 4 cmH
2
O.
2.
Switch to
PSV
mode: as VT is not used in this mode, the setting is not applied but it remains
saved to the memory. However, the PEEP is applied.
3.
Setting of the PEEP at 6 cmH
2
O.
4.
Switch to
(A)VCV
mode: the PEEP setting is saved. The VT setting is applied again.