The respiratory rate can be increased as soon as the ventilator detects that the patient is making a res-
piratory effort
Set-points
FiO2
inspired oxygen fraction
VT
tidal volume (mL)
RR
minimum respiratory
frequency (bpm)
PEEP
positive end of expiration pressure
I:E
ratio of inspiration time to expiration time
I. Trig
inspiratory trigger (L/min)
Tplat
adjustment of inspiratory plateau time (% TI)
Flow shape
form of flow rate insufflated to the patient: Constant or decelerated
VT sigh
enabling the sigh function if ≠ OFF
sigh amplitude (unit x VT; e.g. VT sigh = 1.5 VT)
Sigh
sigh periodicity (1 sigh every x cycles)
Note:
In VCV, there is no apnea ventilation. Safety is guaranteed by setting the lower limit on
the frequency and tidal volume according to each patient category.
4.6.3 (A)PCV (controlled ventilation or assisted pressure-controlled ventilation)
[PCV = Pressure-Controlled Ventilation]
Principle
PCV mode is used to control the pressure delivered to the patient, the inspiration time, and the fre-
quency of the ventilation cycles. During expiration, the ventilator regulates the pressure in order to
maintain the set PEEP level.
If the inspiratory trigger value is set, this mode denotes (A) to show that it is now also assisted.
The respiratory rate can be increased as soon as the ventilator detects that the patient is making a res-
piratory effort.
Set-points
FiO2
inspired oxygen fraction
PI
inspiratory pressure (cmH
2
O)
RR
minimum respiratory rate (bpm)
PEEP
positive end of expiration pressure (cmH
2
O)
I:E
ratio of inspiration time to expiration time
I. Trig
inspiratory trigger (L/min)