Monnal T60 Advanced - User manual
It is also necessary to carry out the automatic tests with the combination of all of the components of
the ventilator's ventilation system foreseen for the patient and to obtain faultless test results.
Patient circuit
Adult single-use patient circuit with water trap
Catalog Number
KG019300
Resistance at 60 L/min
1.2 cmH
2
O
Compliance
1.2 mL/cmH
2
O
Volume
1050 cm3
Adult single-use patient
Catalog Number
KG020100
Resistance at 60 L/min
0.4 cmH
2
O
Compliance
1 ml/cmH
2
O
Volume
1216 cm3
Pediatric single-use patient circuit with water trap
Catalog Number
KG019400
Resistance at 30 L/min
5 cmH
2
O
Compliance
0.6 mL/cmH
2
O
Volume
760 cm3
Pediatric single-use patient circuit without water trap
Catalogue Number
KG020200
Resistance at 5 L/min
0.32 cmH
2
O
Compliance
0.8 ml/cmH
2
O
Volume
530 cm3
Bacteriological filter
Ventilator outlet bacteriological filter
Catalogue Number
KV103300
Resistance at 60 L/min
1.7 cmH
2
O
Compliance
0.1 mL/cmH
2
O
Volume
120 cm3