Monnal T60 Advanced - User manual
The ventilation screen in the ECLS environment
1
Launch of the
Default (A)VCV
mode and
deactivation of the ECLS environment.
2
Display of the ventilation mode in the ECLS
environment.
3
Settings available in ECLS for this mode.
4
Alarm thresholds in ECLS for this mode.
Deactivating the ECLS environment
To continue ventilation outside the ECLS environment, and no longer adjust the alarm thresholds to
the ECMO context, a
Default (A)VCV
mode can be launched.
Press
(A)VCV
(1)
. Validate with the rotating knob.
Note:
The ECLS environment may also be deactivated by selecting another ventilation mode.
In this case, the ventilation mode is launched without the ECLS environment.