5 Alarms and other messages
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
142 Low spontaneous RR!!
MEDIUM
Measured spontaneous
respiratory rate lower
than the alarm threshold
Inhibition for 60 s when
ventilation starts.
Inhibition during a
pause and 30 s after the
pause.
Triggering after 30 s
Check the consistency
between the alarm lev-
els and the ventilation
settings.
143 High spontaneous RR!!
MEDIUM
Measured spontaneous
respiratory rate higher
than the alarm threshold
Inhibition for 60 s when
ventilation starts.
Inhibition during a
pause and 30 s after the
pause.
Triggering after 30 s
Check the consistency
between the alarm lev-
els and the ventilation
settings.
144 Cstat calculation in
progress
None
(INFO cat-
egory)
Press the
Cstat
mea-
surement key.
Immediate triggering
Not applicable.
Activation delay: Sum of the maximum alarm condition delay and the delay to activation of the audible
or visual alarm
Note:
Delay in the activation of certain alarms can vary according to the status of the ma-
chine (in stand-by, during ventilation, during automatic tests), the type of patient and the
elapsed duration of ventilation.
97