Monnal T60 Advanced - User manual
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
92
O2 mixer defect!!! Venti-
lation Effective to 21%
HIGH
Electronics failure
Immediate triggering
Set the FiO2 set-point =
21%
Re-run the automatic
tests with an O2 source.
Use a different unit and
contact your technical
department
Note: O2 leakage inside
the machine is possible.
93
O2 mixer defect!!! Venti-
lation Effective to 21%
HIGH
Electronics failure
Immediate triggering
Set the FiO2 set-point =
21%
Use a different unit and
contact your technical
department
94
Low oxygen supply
pressure!!!
MEDIUM
O2 supply pressure <
2.8 bars
Triggered within 2 s
Check remaining O2 lev-
el in
cylinder or pressure in
supply network.
For operation on a low-
pressure source, select
‘Low-pressure O2’ from
the configuration menu.
95
O2 maximum for 2 min!
None
(INFO cat-
egory)
Press the “100% O2” key Start of next ventilation
cycle
Not applicable.
96
O2 low pressure
None
(INFO cat-
egory)
Press the “Low-pressure
O2” key
Immediate triggering
Not applicable.
97
O2 mixer defect!!! Safe-
ty Ventilation
HIGH
Electronics failure
Immediate triggering
Use a different unit and
contact your technical
department
98
No internal battery!!!
HIGH
No internal battery or in-
ternal battery fully dis-
charged
Immediate triggering
Use a different unit and
contact your technical
department.
The alarm is active un-
til the internal battery
starts charging.
99
Extractable battery is
missing!!
MEDIUM
Absence of the ex-
tractable battery with
the
Extractable bat-
tery alarm
option acti-
vated.
Immediate triggering
Replace the extractable
battery or deactivate the
extractable battery op-
tion.
92