Markings
Autoclavable
Single-use
serial number
product reference
Expiratory flow sensor
product reference
batch number
Single-use version
The expiratory assembly is delivered by
Air Liquide Medical Systems
clean and ready for clinical use.
The exchange interval depends on hospital infection-control protocols, the presence of secretions and
the nebulization of drugs.
CAUTION:
Reusing single-use accessories or consumables carries the risk of patient cross-
contamination.
Autoclavable version
To help track the number of cycles, the sensor and valve are marked with a unit serial number (SN).
CAUTION:
The expiratory flow sensor requires special precautions during pre-disinfection,
cleaning, and disinfection. It contains a very fine, fragile platinum wire.
It is therefore important to:
•
Avoid inserting any object into the flow sensor,
•
Avoid exposing it to a jet of water or air,
•
Avoid any impacts or dropping it.
After a cleaning/disinfection cycle (immersion, autoclave, etc.), the hot wire expiratory flow sensor
must be thoroughly dried before being reinstalled, connected to the expiratory valve, and connected to
the ventilator.
Perform a visual check to ensure that the various components are in good condition.
CAUTION:
Sterilization of the autoclavable expiratory valve flow sensor and autoclavable ex-
piratory valve before the first use is recommended.
Expiratory assembly cleaning protocol
See Expiratory assembly cleaning protocol on page 141.