Monnal T60 Advanced - User manual
Curves
Volume (ml)
Adjustable on successive scales: 0 to 100, 0 to 500, 0 to 1000, 0 to
3000
Instantaneous CO2 / Av-
erage CO2 concentration
(mmHg)
Adjustable on successive scales:
0 to +50, 0 to +100
Time (s)
Adjustable on successive scales:
0 to +6, 0 to +12, 0 to +18
Loop curves
P/V, D/P, D/V and V/CO2
Data storage
Alarm log
List of alarms triggered during the use of the unit
Trends
Simultaneous display of two parameters measured during a maximum period
of 80 h.
All measured parameters are accessible
9.2.1.8 Ventilation specifications
Ventilation modes
VCV (controlled ventilation or assisted vol-
ume-controlled ventilation)
(A)VCV
PCV (controlled ventilation or assisted pres-
sure-controlled ventilation)
(A) (A)PCV
PSV (spontaneous ventilation with inspiratory as-
sistance and PEEP)
PSV
CPAP (Continuous Positive Airway Pressure)
CPAP
Synchronised Intermittent Mandatory Ventilation VSIMV
Airway Pressure Release Ventilation
APRV
Synchronized Intermittent Mandatory Pressure
Monitored Ventilation
PSIMV
Pressure-regulated volume-controlled ventilation PRVC
Spontaneous ventilation with inspiratory assis-
tance, PEEP and servomechanism frequency
PS-Pro
22
The unit records a maximum of 4000 events. An event is represented by a set-point or an alarm
threshold when an adjustment is made and an alarm if triggered.
118
Summary of Contents for MONNAL T60 ADVANCED
Page 2: ......