Monnal T60 Advanced - User manual
Note:
When the log reaches its maximum size, the oldest events will be overwritten by in-
coming events.
5.7 Table of alarms
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
1
Stop ventilation?
MEDIUM
Device switched to
Stand-by Mode
Immediate triggering
Press the control wheel
to stop ventilation.
Press ‘Cancel’ to main-
tain
ventilation.
2
Ventilator shutdown?
MEDIUM
Unit Off request
Immediate triggering
Press the control wheel
to shut down the unit.
Press ‘Cancel’ to leave
the unit in Standby.
3
Expiratory branch may
be obstructed!!!
HIGH
Patient circuit obstruct-
ed, or expiratory valve
membrane stuck, or an
electronic fault has oc-
curred
Triggering after two ven-
tilation cycles with high
pressure and PEEP + 5
cmH2O or
PEEP + 5 cmH2O for 15
consecutive seconds
Check the patient cir-
cuit.
Check the expiratory
valve membrane.
Contact your technical
department if the prob-
lem persists
Expiration is prolonged
for a maximum of 15
s, until the measured
PEEP returns below the
PEEP setting.
4
Ventilation Interrupt-
ed!!!use an alternative
ventilator
HIGH (UL-
TRA cate-
gory)
Electronics failure
Immediate triggering
Use a different unit and
contact your technical
department
Patient circuit vented to
atmosphere
5
Ventilation Interrupted!!!
use an alternative venti-
lator
HIGH (UL-
TRA cate-
gory)
Electronics failure
Immediate triggering
Use a different unit and
contact your technical
department
Patient circuit vented to
atmosphere
6
Ventilation Interrupted!!!
use an alternative venti-
lator
HIGH (UL-
TRA cate-
gory)
Electronics failure
Immediate triggering
Use a different unit and
contact your technical
department
Patient circuit vented to
atmosphere
7
Safety Ventilation!!! Rec-
ommend switching ven-
tilator
HIGH
Electronics failure
Immediate triggering
Use a different unit and
contact your technical
department
82