Monnal T60 Advanced - User manual
It requires the use of a humidifier and nasal cannula (or tracheotomy tube) specific to high throughput
oxygen therapy.
Note:
Please refer to the patient/machine interface manual.
Set-up
1.
Install the humidifier on the
Monnal T60 Advanced
stand;
Note:
Place the humidifier lower than the patient to prevent water flowing into the cir-
cuit.
2.
Perform an automatic test with a dual branch circuit (see section Automatic tests on page 25);
3.
Disconnect the dual branch circuit and set up the required patient circuit on the inspiratory
branch of the ventilator and on the humidifier;
4.
Connect the patient/machine interface between the humidifier and the patient;
5.
Check that the humidifier is in working order before switching it on;
6.
Activate the Hflow function with the desired flow-rate setting.
Note:
Humidify the flow generated by the device.
CAUTION:
The Hflow function is not a ventilation mode: it is therefore necessary to monitor
its use. The patient breathes in a fully autonomous manner, this is why:
•
Apnea ventilation is not ensured in the event of a patient’s respiratory arrest.
•
No alarm is set off if the patient circuit is disconnected.
CAUTION:
Monitored FiO
2
is the FiO
2
delivered by the ventilator. If the patient’s inspiratory
flow is greater than the flow set-point or if the nasal cannula are not properly positioned, then
monitored FiO
2
is potentially higher than that inhaled by the patient.
Note:
The Ppeak threshold can be adjusted. Its default setting is 45 cmH
2
O.
48