9 Technical description
Measured parameters
Mea-
sure-
ment
Reso-
lution
Chest compression rate
(fCC, bpm)
55 -
145
1
±10%
4 s + 5 cycles
Maximum pressure variation
(P-P, cmH2O)
5 - 20
0.1
± 4 cmH2O + 30%
5 cycles
--
--
±2s
-
%CC (%)
0 - 100 1
15%
-
Driving pressure (DP, mL)
0 - 50
1
± (2 cmH2O + 4%)
1 cycle
Expired volume/predicted weight (Vte / PBW,
cmH2O)
0 - 20
1
± 1 cm
-
Spontaneous RR (spont RR, c/min)
1 - 120 1
± 1
4 spontaneous cycles
Spontaneous expired minute volume (spont
MVe, L/min)
0,5 -
40
0.1
± (0.5 L/min + 15%)
30 s
Static compliance (C stat, mL/cmH2O)
10 -
150
0.1
± (5 cmH2O + 20%)
-
Estimated compliance (DynC, mL/cmH2O)
10 -
150
0.1
± (5 cmH2O + 30%)
-
9.2.1.7 Monitoring specifications
Curves
Flow rate (L/min)
Adjustable on successive scales: -10 to +10, -20 to +20, -40 to +40, -80
to +80, -160 to +160
Pressure (cmH
2
O).
Adjustable on successive scales: 0 to +20, 0 to +40, 0 to +60, 0 to +100
14
These accuracy values are ensured in the worst configuration of the ventilation system compris-
ing a bacteriological filter, an adult patient circuit with water trap, an empty humidifier and the
IRMA probe
™
.
15
Filtering is based on a rolling average of n samples.
16
The oxygen sensor complies with oxygen-monitor standards and the following specifications.
17
CO
2
monitoring is performed by the IRMA
™
probe, whose characteristics are specified in section
Specifications of accessories on page 113.
18
In CPV mode only. See CPV: Cardio-Pulmonary Ventilation on page 45.
19
Available in VCV/(A)VCV and VSIMV mode only.
20
Available when the patient's information (size/gender) is entered.
21
In PS Pro and CPAP mode only.
117