9 Technical description
External DC source (ambulance, aeroplane, helicopter)
Type
Rated voltage: 13-24 VDC (tolerance: 11.1-30
VDC)
Permissible voltage: 11.1V to 30V for use with
the
Monnal T60 Advanced
power cable
Maximum current: 11 A
Cable type
2.5m
Monnal T60 Advanced
power cable
(YR123700)
(YR115100) and extractable battery (KY692800)
Type
Lithium-ion
Rated voltage: 22.2 V
Rated capacity: 2600 mAh
Peak current: 5.5 A
Battery capacity (new and charged battery) with
the standard ADULT ventilation configuration
Typically 2.5 hours per battery, or a total of 5
hours (at 25°C ambient temperature
Charging time (h)
Typically 2 hours 20 minutes per battery (quick
charge with machine in stand-by),
Typically 5.5 hours per battery (slow charge with
machine ventilating) according to environmental
conditions and type of power supply
Note:
The internal battery has a life of around 300 charge and discharge cycles or two years,
whichever happens first.
9.2.1.4 Specifications of accessories
IRMA
™
probe
For more information, please refer to the IRMA
™
user manual from MASIMO.
Components of the ventilation system
Components of the ventilator's ventilation system that meet the requirements of the following stan-
dards may be used:
•
ISO 5367 (patient circuits)
•
ISO 23328-1 and ISO 23328-2 (filters)
•
ISO 8185 or ISO 80601-2-74 (humidifiers)
•
ISO 9360-1 or ISO 9360-2 (HME filters)
13
The internal battery is a spare part. For more information, please contact the technical depart-
ment.