![Air Liquide MONNAL T60 ADVANCED User Manual Download Page 86](http://html1.mh-extra.com/html/air-liquide/monnal-t60-advanced/monnal-t60-advanced_user-manual_2875003086.webp)
Monnal T60 Advanced - User manual
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
29
Batteries nearly dis-
charged!!! Connect AC
power.
MEDIUM
The available remaining
time for the internal bat-
tery is low.
Immediate
triggering
Insert a charged, re-
placement auxiliary bat-
tery or connect the unit
to AC mains.
As the internal battery
approaches the end of
its life time, certain ven-
tilation parameters can
be adjusted to optimize
the ventilation time.
31
Battery charger ineffec-
tive!!! Ventilation Effec-
tive
HIGH
The battery charger is
faulty.
Triggering within 10 s
Use a different unit and
contact your technical
department
33
Apnea ventilation!!!
HIGH
No patient respirato-
ry activity for a time
greater than the set T
apnea
T Apnea
Immediate
triggering
Check patient safety
and verify the switch to
back-up ventilation
34
Safety Ventilation!!!
Recommend switching
ventilator
HIGH
Electronics failure
Immediate
triggering
Use a different unit and
contact your technical
department
36
Safety Ventilation!!!
Recommend switching
ventilator
HIGH
Electronics failure
Immediate
triggering
Use a different unit and
contact your technical
department
37
Safety Ventilation!!!
Recommend switching
ventilator
HIGH
Electronics failure
Immediate
triggering
Use a different unit and
contact your technical
department
38
Safety Ventilation!!!
Recommend switching
ventilator
HIGH
Electronics failure
Triggering during auto-
matic tests
Use a different unit and
contact your technical
department
39
Patient circuit leak de-
tected during automatic
tests!!!
HIGH
Patient circuit connec-
tion problem
Triggering on exit from
automatic tests
Check and re-connect
the patient circuit com-
ponents carefully and
repeat the automatic
tests. If the problem
persists, change the pa-
tient circuit.
40
Safety Ventilation!!!
Recommend switching
ventilator
HIGH
Electronics failure
Immediate
triggering
Use a different unit and
contact your technical
department
41
Safety Ventilation!!!
Recommend switching
ventilator
HIGH
Electronics failure
1 s in stand-by
Use a different unit and
contact your technical
department