Note:
A unique serial number on the expiratory flow sensor
(1)
and the valve body
(2)
gives
the date of manufacture of the components and can be used to track the number of cycles
they have undergone.
Reassembly
Reassemble the expiratory assembly wearing gloves: place a new membrane in the valve body.
CAUTION:
It is necessary to perform an automatic test after each assembly of the expiratory
valve and the flow sensor.
CAUTION:
Incorrect reassembly of the expiratory assembly (wrong positioning of the mem-
brane in the valve body, removal or addition of an additional silicon disc, etc.) can lead to de-
graded performance of the device (risk of leaks, incorrect pressurization, failure to respect
the volume – hypoventilation).
10.3.1 Monnal EVA autoclavable expiratory assembly
KY632200 (x1)
KY694500 (x1)
10.3.2 Monnal EVA single-use expiratory assembly
KY664500 (x5)
KY664600 (x20)
KY694800 (x5)
KY694900 (x20)