9 Technical description
Adult
Child
Infant
Description
Modes
Name
Unit
min~max [de-
fault]
res.
min~max [de-
fault]
res.
min~max [de-
fault]
res.
Peak flow
(A)VCV,
(A)VCV
NIV,
VSIMV
Flow
rate
L/min
4~150 [28]
1
4~150 [15]
1
2~150 [4]
1
TI/Ttot
(A)VCV,
(A)VCV
NIV,
(A)PCV,
(A)PCV
NIV
TI/Ttot
%
10~50 [33]
1
10~50 [33]
1
10~50 [33]
1
Apnea ventilation
volume
All ex-
cept
(A)VCV,
(A)VCV
NIV,
(A)PCV,
(A)PCV
NIV,
PS-Pro,
PRVC,
APRV,
CPV
VT ap-
nea
mL
100~2,000 [480]
10
50~500 [120]
5
20~75 [35]
5
Apnea ventilation
frequency
All ex-
cept
(A)VCV,
(A)VCV
NIV,
(A)PCV,
(A)PCV
NIV,
PS-Pro,
PRVC,
APRV,
CPV
RR ap-
nea
c/min
5~40 [15]
1
5~60 [25]
1
10~80 [40]
1
Trigger duration
All ex-
cept
(A)VCV,
(A)VCV
NIV,
(A)PCV,
(A)PCV
NIV,
PS-Pro,
PRVC,
APRV
T ap-
nea
s
15~60 [20]
1
4~60 [20]
1
2~60 [10]
1
Trigger duration
CPAP,
CPAP
NIV
T ap-
nea
s
OFF, 15~60 [20]
1
OFF, 4~60 [20]
1
OFF, 2~60 [10]
1
Size
Setting
Size
cm
106~235 [--]
1
73~166 [--]
1
45~77 [--]
1
Size
Setting
Size
Inches
42~92 [--]
1
29~65 [--]
1
17~30 [--]
1
Gender
Setting
Gender
NA
F~M [--]
F~M [--]
F~M [--]
25
PI = PS + PEEP
26
Setting based on increments of 4 between 21 and 25, then 5 as from 25.
27
Setting based on increments of 0.05 between 0.25 and 0.3, then 0.1 as from 0.3.
28
Fixed TI of 0.8s.
129
Summary of Contents for MONNAL T60 ADVANCED
Page 2: ......
Page 140: ...Monnal T60 Advanced User manual 10 2 Maintenance data sheet 140...