Monnal T60 Advanced - User manual
High pressure (HP) and low pressure (LP) O
2
inlets
Standalone mode
Cylinder (HP)
Low flow O2 metered supply e.g. Concentrator
(LP)
Mixer
Electronic, electrochemical O2 sensor
Precision (% of set value)
± 5 percentage points under the conditions spec-
ified by standard, ISO 80601-2-12
Gas consumption
Patient venti 4 L/min
(flow-by + internal consumption)
Connectors
Inspiratory hose connector
ISO 22 mm male
Expiratory hose connector
ISO 22 mm male
Inspiratory and expiratory resistances
Resistances at 60 L/min (cmH
2
O)
(m single-use adult patient circuit
K filter KV103300)
Inspiratory: 4.73
Expiratory: 4.89
Resistances at 30 L/min (cmH
2
O)
(m single-use child patient circuit
K filter KV103300)
Inspiratory: 2.62
Expiratory: 2.52
Resistances at 5 L/min (cmH
2
O)
(m single-use child patient circuit
KG filter KV103300)
Inspiratory: 0.28
Expiratory: 1.39
Pressure
Maximum limited pressure (Plimmax)
90 cmH
2
O: blower performance limitation
Maximum working pressure (P w max)
70 cmH
2
O
Minimum working pressure (P w min)
0 cmH
2
O
Minimum limited pressure (P lim min)
Back-up ambient air intake preventing a pressure
drop in the patient circuit
23
Consumption example:
120