Monnal T60 Advanced - User manual
Particular case of peak airway pressure (Ppeak) and maximum tidal volume (MTV)
In all modes, exceeding the upper Ppeak threshold switches off the current inspiration and is forced
to change to expiration. Furthermore, in non-invasive pressure-controlled ventilation modes, exceed-
ing the upper insufflated tidal volume (VTi) threshold also switches off the current inspiration and is
forced to change to expiration.
4.6.17.2 Automatic thresholds
The
Auto. set
button is used to automatically configure all of the alarm thresholds according to the
values measured at the time when the button is pressed (see Adjustment of alarm thresholds on page
57).
CAUTION:
After pressing the
Auto. Set
button, check that the thresholds obtained in this way
are appropriate for the clinical condition of the patient.
58